首页 | 本学科首页   官方微博 | 高级检索  
     


Altered phase-encoding order for reduced sensitivity to motion in three-dimensional MR imaging.
Authors:H W Korin  S J Riederer  A E Bampton  R L Ehman
Affiliation:Magnetic Resonance Laboratory, Mayo Clinic and Foundation, Rochester, MN 55905.
Abstract:
A method of reordering phase and slab encoding that can be used to address some of the inherent problems due to motion in three dimensional imaging is described and implemented. The method is shown to be more robust with respect to reducing artifacts resulting from several fundamental types of motion. It can be readily implemented on a standard magnetic resonance imager with essentially no increase in total imaging time. Results of simulations and phantom and in vivo experiments are presented.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏