Automatic Atrial Anti-Tachy Pacing for the Termination of Spontaneous Atrial Tachyarrhythmias: Clinical Experience with a Novel Dual-Chamber Pacemaker |
| |
Authors: | Dirk Vollmann Jens Stevens Arnd B. Buchwald Christina Unterberg |
| |
Affiliation: | (1) Department of Cardiology and Pneumology, Georg-August-University, Goettingen, Germany |
| |
Abstract: | Automatic atrial anti-tachy pacing (aATP) is a novel approach to treat paroxysmal/persistent atrial tachyarrhythmias in pacemaker patients. To evaluate the efficacy of aATP in terminating spontaneous atrial flutter/tachycardia episodes (AT), a dual-chamber stimulator with extensive diagnostic capabilities and programmable aATP therapies (AT500TM, Medtronic Inc.) was implanted in 30 patients with conventional pacing indications. During a mean follow-up time of 5.5 (1–12) months, aATP was delivered for 10494 AT. According to automatic device analysis, 8289 AT were treated with success (success-rate 79.0%). On 468 AT stored with the corresponding atrial EGM, an additional manual analysis was performed. The success-rate based on automatic analysis of these AT episodes (73.1%) was comparable to that found for all treated AT (79.0%), but manual EGM analysis revealed that only 209 of the 468 treated AT episodes (44.7%) were actually terminated by aATP. The aATP success-rate in the slower (cycle length 360–270ms) AT detection zone was significantly higher (73.8%, 62/84eps) than in the overlapping, faster (cycle length 270–220ms) AT zone (38.3%, 147/384eps, P<0.01).Conclusions: According to manual analysis, 1. aATP was safe and had a success-rate of 44.7%, 2. aATP success-rate was higher for AT in the slower than in the faster detection zone and 3. automatic analysis overestimated the efficacy of aATP. |
| |
Keywords: | atrial tachyarrhythmias dual-chamber pacemaker atrial flutter anti-tachy pacing success-rate |
本文献已被 SpringerLink 等数据库收录! |
|