首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of gaseous microemboli counts in arterial, simultaneous and venous heat exchange with a hollow fiber membrane oxygenator
Authors:Sutton R G  Riley J B  Merrill J H
Affiliation:Medical University of South Carolina College of Health Related Professions, Extracorporeal Circulation Technology Department, Charleston.
Abstract:
Potential sources of gaseous microemboli during cardiopulmonary bypass are varied. However, it is known that membrane oxygenators generate fewer gaseous microemboli than bubble oxygenators and that bubblers cannot utilize arterial heat exchange without generating significant gaseous microemboli during rewarming. A membrane oxygenator utilizing simultaneous gas and heat exchange raises the concern that concurrent gas and heat exchange would result in a higher production of gaseous microemboli compared to conventional venous heat exchange devices. This in vitro study compared venous, simultaneous, arterial and control (venous) heat exchanger gaseous microemboli counts during rewarming. No significant difference was found between the four heat exchangers when comparing inlet and outlet gaseous microemboli counts. This in vitro study suggests that there is no difference in gaseous microemboli generation when varying the position of the heat exchanger in the extracorporeal circuit incorporating a microporous membrane oxygenator.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号