High-performance liquid chromatographic assay for the determination of nilotinib in human plasma |
| |
Authors: | Yuki Misato Yamakawa Yuji Uchida Takashi Nambu Takeru Kawaguchi Tatsuya Hamada Akinobu Saito Hideyuki |
| |
Affiliation: | Department of Clinical Pharmaceutical Sciences, School of Pharmaceutical Sciences, Kumamoto University, 5–1 Oehonmachi, Kumamoto 862–0973, Japan. |
| |
Abstract: | A precise and convenient high-performance liquid chromatography (HPLC) method has been established to assay nilotinib in human plasma. Chromatographic separation of nilotinib was performed on a LiChrosphere(?)100 RP-18(e) column (250 mm×4.0 mm, 5 μm) using a mixture of acetonitrile and 0.01 M phosphate buffer (pH 3.0) (42 : 58, v/v) under isocratic conditions at a flow rate of 1.0 ml/min with ultraviolet (UV) detection at 266 nm. The calibration curve showed linearity at concentrations between 250 ng/ml and 5000 ng/ml (r(2)>0.999). The mean±S.D. absolute recovery of nilotinib from plasma was 99.2±3.3%. The coefficients of variation of both intra- and inter-day precision were below 9.1%. These results indicate that this new HPLC-based quantification may be useful for therapeutic drug monitoring of nilotinib to help manage treatment in patients with chronic myeloid leukemia in clinical practice. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|