首页 | 本学科首页   官方微博 | 高级检索  
     


Interventional MR imaging at 1.5 T: quantification of sound exposure
Authors:Moelker Adriaan  Maas Ronald A J J  Lethimonnier Franck  Pattynama Peter M T
Affiliation:Department of Radiology, Erasmus University Medical Center Rotterdam, 50 Dr Molewaterplein, P.O. Box 1738, the Netherlands. moelker@rdiag.fgg.eur.nl
Abstract:
Sound pressure levels (SPLs) during interventional magnetic resonance (MR) imaging may create an occupational hazard for the interventional radiologist (ie, the potential risk of hearing impairment). Therefore, A-weighted and linear continuous-equivalent SPLs were measured at the entrance of a 1.5-T MR imager during cardiovascular and real-time pulse sequences. The SPLs ranged from 81.5 to 99.3 dB (A-weighted scale), and frequencies were from 1 to 3 kHz. SPLs for the interventional radiologist exceeded a safe SPL of 80 dB (A-weighted scale) for all sequences; therefore, hearing protection is recommended.