10-Year Follow-Up Wear Analysis of First-Generation Highly Crosslinked Polyethylene in Primary Total Hip Arthroplasty
Authors:Nimrod Snir  Ian D. Kaye  Christopher S. Klifto  Mathew J. Hamula  Theodore S. Wolfson  Ran Schwarzkopf  Fredrick F. Jaffe
Affiliation:1. Department of Orthopaedic Surgery, New York University Hospital for Joint Diseases, New York, New York;2. Department of Orthopaedic Surgery, University of California Irvine Medical Center, Orange, California
Abstract:
Our goal was to report a 10-year follow up of linear penetration rates for HCLPE, and to determine whether a difference exists between penetrations measured on pelvis or hip anterior–posterior radiographs. We reviewed 48 total hip arthroplasties where a first-generation HCLPE liner was used. Femoral head penetration was measured on both AP pelvis and hip radiographs. Total wear and wear rate at 10 years were 1.26 mm and 0.122 mm/y, respectively. The rate decreased significantly after the first 2–3 years, plateauing at a wear rate of 0.05 mm/y for the last 5 years. The AP hip total wear and wear rate were 1.38 mm and 0.133 mm/y respectively, while rates were 1.13 mm and 0.109 mm/y respectively for the pelvis radiographs (P < .05). We found a significant difference in measurements of linear penetration when comparing AP pelvis vs. hip radiographs with lower rates recorded using an AP pelvis.
Keywords:total hip arthroplasty   highly cross linked polyethylene   wear analysis   first generation   10-year follow-up
本文献已被 ScienceDirect 等数据库收录!