宁夏2018参加全国个人剂量监测系统盲样比对结果分析
引用本文:姬绪莉,孔庆宇,卢桂才,孙彦玲,李鸿成. 宁夏2018参加全国个人剂量监测系统盲样比对结果分析[J]. 中国辐射卫生, 2020, 29(2): 182-184. DOI: 10.13491/j.issn.1004-714X.2020.02.021
作者姓名:姬绪莉  孔庆宇  卢桂才  孙彦玲  李鸿成
作者单位:宁夏疾病预防控制中心,宁夏 银川 750004
摘    要:
目的 通过参加全国放射工作人员个人剂量监测系统盲样比对,了解本省实验室个人剂量监测系统运行状况,提高监测能力和水平。方法 按照《2018年度全国个人剂量监测能力考核方案》要求准备热释光剂量计,邮寄回考核组盲样照射后,由参加比对实验室检测并进行数据分析。结果 本次5组比对结果单组性能偏差在-0.07~0.02(均符合|Pi| ≤ 0.10),同时综合性能偏差B2+S2 ≤ 0.30,故而实验得70满分,因检测报告得9分,总分未达到评定优秀的要求,本实验室个人剂量监测系统盲样比对结果为合格。结论 宁夏本实验室个人剂量监测系统运行良好,检测结果准确、误差范围较小,可满足国家职业性外照射个人剂量监测相关标准要求。

关 键 词:个人剂量  盲样考核  单组性能  综合性能  
收稿时间:2019-08-21

Analysis on the result of the 2018 national individual does monitoring system blind sample comparison in Ningxia
JI Xuli,KONG Qingyu,LU Guicai,SUN Yanling,LI Hongcheng. Analysis on the result of the 2018 national individual does monitoring system blind sample comparison in Ningxia[J]. Chinese Journal of Radiological Health, 2020, 29(2): 182-184. DOI: 10.13491/j.issn.1004-714X.2020.02.021
Authors:JI Xuli  KONG Qingyu  LU Guicai  SUN Yanling  LI Hongcheng
Affiliation:Ningxia Center for Disease Control and Prevention, Yinchuan 750004 China
Abstract:
Objective Through participating in the blind sample comparison of the national personal dose monitoring system for radiation workers, to understand the operation status of personal dose monitoring system of the provincial laboratory, and improve the monitoring ability and level.Methods According to the requirements of the national personal dose monitoring ability assessment program in 2018, to prepare the thermoluminescent dosimeter, mail them back to the assessment group for blind sample irradiation, and then participate in the comparison laboratory inspection test and analyze the data.Results In the comparison results of the 5 groups, the performance deviation of single group in 5 groups of comparison results is - 0.07~0.02 (all in line with |Pi| = 0.10), and the comprehensive performance deviation is B2 + S2 ≤ 0.30, so the experiment got a full score of 70. As the test report got 9 points, the total score did not meet the requirements of excellent evaluation, the blind sample comparison result of personal dose monitoring system in this laboratory is qualified.Conclusion The personal dose monitoring system of our laboratory in Ningxia is in good operation, the test results are accurate and the error range is small, and which can meet the relevant national standards for personal dose monitoring of occupational external exposure.
Keywords:Personal Dose  Blind Sample Test  Single Group Performance  Comprehensive Performance  