首页 | 本学科首页   官方微博 | 高级检索  
     


A comparison of the outcome of cemented all-polyethylene and cementless metal-backed acetabular sockets in primary total hip arthroplasty
Authors:Hartofilakidis George  Georgiades George  Babis George C
Affiliation:University of Athens Medical School, Athens, Greece.
Abstract:
We compared, after a 10-year-minimum follow-up, the outcome of 50 cemented all-polyethylene Charnley acetabular sockets with that of 51 cementless metal-backed sockets in 2 comparable cohorts of young patients. Although the revision rate for the cemented and cementless group was 28% and 35%, respectively, the revision rate for aseptic loosening was 28% for the cemented and 12% for the cementless group. The mean polyethylene wear was 0.112 and 0.114 mm/y, respectively, for the 2 groups. Linear osteolysis was observed in 18 of 50 cemented sockets. Expansile osteolysis presented in 10 of 51 cementless sockets and only in one of the cemented sockets. In conclusion, cementless components had more durable fixation than cemented components. However, they presented more aggressive expansile osteolysis caused by the coexistence of polyethylene and metal debris.
Keywords:total hip arthroplasty   acetabular fixation   cemented   cementless
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏