High-definition freehand 3-D ultrasound |
| |
Authors: | Treece Graham M Gee Andrew H Prager Richard W Cash Charlotte J C Berman Laurence H |
| |
Affiliation: | Department of Engineering, Trumpington Street, Cambridge, UK. gmt11@eng.cam.ac.uk |
| |
Abstract: | This paper describes a high-definition freehand 3-D ultrasound (US) system, with accuracy surpassing that of previously documented systems. 3-D point location accuracy within a US data set can be achieved to within 0.5 mm. Such accuracy is possible through a series of novel system-design and calibration techniques. The accuracy is quantified using a purpose-built tissue-mimicking phantom, designed to create realistic clinical conditions without compromising the accuracy of the measurement procedure. The paper includes a thorough discussion of the various ways of measuring system accuracy and their relative merits; and compares, in this context, all recently documented freehand 3-D US systems. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|