首页 | 本学科首页   官方微博 | 高级检索  
     

尿碘测定中质量控制图的应用探讨
引用本文:陈春祝,谢维平,欧阳燕玲. 尿碘测定中质量控制图的应用探讨[J]. 中国地方病学杂志, 2011, 30(3). DOI: 10.3760/cma.j.issn.1000-4955.2011.03.033
作者姓名:陈春祝  谢维平  欧阳燕玲
作者单位:福建省泉州市疾病预防控制中心理化检验科,362000
摘    要:
目的 针对现行尿碘检测方法中测定误差因素多的特点,建立实验室内质量控制体系,确保检验数据的准确可靠.方法 收集日常检测中空白管吸光度值及标准物质检测值,计算均值、标准偏差、控制限、辅助线,以它们为线绘制空白管吸光度值控制图及相对误差控制图;用不同碘含量(高、中、低浓度)的试样测定结果,计算各组平均值、相对减差值、加权值、临界限(Rc)值,列出临界限Rc值控制表.结果 空白值吸光度控制限((x)±3sb)范围为1.183~1.553;精密度控制临界限Rc值在尿碘0-<150 μg/L时为0.0883,150~300 μg/L时为0.0572;相对误差控制限(E+3se)为9.3%.结论 建立质量控制图可有效控制实验室内尿碘检测质量,保证尿碘测定结果的可靠性和真实性.
Abstract:
Objective Establish a laboratory quality control system to ensure accurate and reliable test data and to contrapose the influence of error factors in current detection methods for urinary iodine measurement. Methods The results of reagent blank absorbance value and uric iodine standard materials were collected, then their relevant technical indexes such as mean, standard deviation, control limit, auxiliary line were worked out. Then the quality control chart of blank test and the relative error control chart were made base on these technical indexes. And different iodine concentrations (high, middle and low concentration) were tested and their mean,relative reduction difference value, weighted mean value and critical limit Rc value were calculated, and then critical limit Rc value precision control chart was made. Results The range of absorbance of blank control test was 1.183 to 1.553. And the limit of the accuracy control Rc value was 0.0883, 0.0572, respectively, when the concentrations of urinary iodine was 0~ < 150 μg/L and 150 ~ 300 μg/L. The control limit of the relative error was 9.3%. Conclusions The method of quality control chart could be satisfactorily applied to identify the quality of the analytical results of urine iodine, and ensure the results of the urine iodine reliable and authentic.

关 键 词:尿    实验室  质量控制

Application of quality control chart in the determination of urinary iodine
CHEN Chun-zhu,XIE Wei-ping,OUYANG Yan-ling. Application of quality control chart in the determination of urinary iodine[J]. Chinese Jouranl of Endemiology, 2011, 30(3). DOI: 10.3760/cma.j.issn.1000-4955.2011.03.033
Authors:CHEN Chun-zhu  XIE Wei-ping  OUYANG Yan-ling
Abstract:
Objective Establish a laboratory quality control system to ensure accurate and reliable test data and to contrapose the influence of error factors in current detection methods for urinary iodine measurement. Methods The results of reagent blank absorbance value and uric iodine standard materials were collected, then their relevant technical indexes such as mean, standard deviation, control limit, auxiliary line were worked out. Then the quality control chart of blank test and the relative error control chart were made base on these technical indexes. And different iodine concentrations (high, middle and low concentration) were tested and their mean,relative reduction difference value, weighted mean value and critical limit Rc value were calculated, and then critical limit Rc value precision control chart was made. Results The range of absorbance of blank control test was 1.183 to 1.553. And the limit of the accuracy control Rc value was 0.0883, 0.0572, respectively, when the concentrations of urinary iodine was 0~ < 150 μg/L and 150 ~ 300 μg/L. The control limit of the relative error was 9.3%. Conclusions The method of quality control chart could be satisfactorily applied to identify the quality of the analytical results of urine iodine, and ensure the results of the urine iodine reliable and authentic.
Keywords:Urine  Iodine  Laboratories  Quality control
本文献已被 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏