Assessment of cardiac output in children: a comparison between the pressure recording analytical method and Doppler echocardiography. |
| |
Authors: | Marco Calamandrei Lorenzo Mirabile Stefania Muschetta Gian Franco Gensini Luciano De Simone Salvatore M Romano |
| |
Affiliation: | Department of Anaesthesia and Intensive Care, Meyer Children's Hospital, Florence, Italy. m.calamandrei@meyer.it |
| |
Abstract: | OBJECTIVE: To assess cardiac output in pediatric patients with the pressure recording analytical method (PRAM) and the Doppler echocardiography method. PRAM derives cardiac output from beat-by-beat analysis of the arterial pressure profile (systolic and diastolic phase) in the time domain. DESIGN: A prospective observational study. SETTING: Pediatric intensive care unit at a tertiary care children's hospital. PATIENTS: Forty-eight patients between the ages of 1 month and 18 yrs. INTERVENTIONS: Femoral or radial artery catheterization and mechanical ventilation. MEASUREMENTS AND MAIN RESULTS: Cardiac output was simultaneously estimated by Doppler echocardiography and PRAM. Cardiac output values obtained by Doppler echocardiography (2.7 +/- 1.6 L/min, range 0.92-8.20) were significantly correlated with those estimated by PRAM (2.6 +/- 1.7 L/min, range 0.89-7.48; r2 = .99, p < .01). The mean difference between the two estimates was 0.12 +/- 0.27 L x min(-1) (95% confidence interval, -0.54 to 0.77 L x min(-1)). CONCLUSIONS: In the range of ages evaluated, PRAM provides reliable estimates of cardiac output when compared with noninvasive techniques. |
| |
Keywords: | |
|
|