首页 | 本学科首页   官方微博 | 高级检索  
     


Dual Chamber Pacing with a Single Lead System: Initial Clinical Results
Authors:JEAN LUC REY  FOUAD EL GHELBAZOURI  CHRISTOPHE TRIBOUILLOY
Abstract:
A new mode of biphasic pacing was used in 26 patients to assess the feasibility of atrial pacing by means of the floating atrial ring electrodes of a single lead VDD permanent pacing system. During implantation, atrial pacing was possible in 25 patients with a 1-ms total pulse duration, a mean atrial threshold of 1.70 ± 0.60 V (range, 0.6–3.0), and a mean diaphragmatic threshold of 6.7 ± 2.5 V (range, 2.5–10.0). At 3 months, the atrial threshold had increased beyond 4.8 V in three patients. In the 22 other patients, the mean atrial threshold was 2.2 ± 0.5 V (range, 1.50–3.50) in the supine position and 2.5 ± 0.8 V (range, 1.5–4.8) in the sitting position. Stable atrial capture without diaphragmatic stimulation was achieved in 76% of patients.
Keywords:single lead    atrial floating electrodes    atrial pacing    biphasic pacing
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号