The biomechanics of point contact-dynamic compression plate and its effects on bone perfusion
引用本文:赵玉峰 李起鸿 顾祖超 王爱民. The biomechanics of point contact-dynamic compression plate and its effects on bone perfusion[J]. 中华创伤杂志(英文版), 2006, 9(3): 161-167
作者姓名:赵玉峰 李起鸿 顾祖超 王爱民
作者单位:Trauma Division Ⅱ Daping Hospital,the Third Military Medical University,Chongqing 400042,China,Department of Orthopaedics,Southwest Hospital,the Third Military Medical University,Chongqing 400038,China,Department of Orthopaedics,First People's Hospital of Chengdu,Sichuan 510000,China,Trauma Division Ⅱ,Daping Hospital,the Third Military Medical University,Chongqing 400042,China
摘    要:Objective: To compare the mechanical properties of point contact-dynamic compression plate (PC-DCP) and its effects on cortical bone perfusion with that of dynamic compression plates (DCP) in goat tibiae. Methods: Twenty pairs of matched fresh goat tibiae were used. A transverse fracture model was established. The fractures with a 3mm interspace between the fracture ends were subject to fixations with the DCPs and the PC-DCPs respectively, then the four-points bending tests and the torsion tests were conducted to compare the mechanical properties of the PC-DCP with that of DCP. Another 13 sexually mature goats underwent fixations with the DCPs and the PC-DCPs, respectively, at the mid-shafts of the intact bilateral tibiae. Ischemic zones were observed at four time points (1 day ,2,6, and 12 weeks after operation) using disulphine blue staining technique. Results: There were no significant differences in mechanical properties, such as bend- and torsion-resistance, between the DCPs and the PC-DCPs. One day, 2, and 6 weeks after operation, on the side of DCP fixation, outer cortical bone ischemia under the plate persisted, and this condition did not reverse until 12 weeks after operation. However, on the side of PC-DCP fixation, cortical bone ischemia occurred only in the periphery of the screw holes and at the contact sites of the PC NUTs 1 day after operation, and it disappeared at 2 weeks after operation. Conclusions: The PC-DCP has similar biomechanical properties of the DCP, byt is less detrimental to local bone blood circulation than the conventional plates.

关 键 词:生物力学 动力学 骨灌注 病理
收稿时间:2005-04-18

The biomechanics of point contact-dynamic compression plate and its effects on bone perfusion
ZHAO Yu-feng,LI Qi-hong,GU Zu-chao,WANG Ai-min. The biomechanics of point contact-dynamic compression plate and its effects on bone perfusion[J]. Chinese journal of traumatology, 2006, 9(3): 161-167
Authors:ZHAO Yu-feng  LI Qi-hong  GU Zu-chao  WANG Ai-min
Affiliation:1. Trauma Division Ⅱ, Daping Hospital, the Third Military Medical University, Chongqing 400042, China
2. Department of Orthopaedics, Southwest Hospital, the Third Military Medical University, Chongqing 400038, China
3. Department of Orthopaedics, First People's Hospital of Chengdu, Sichuan 510000, China
Abstract:Objective: To compare the mechanical properties of point contact-dynamic compression plate (PC-DCP) and its effects on cortical bone perfusion with that of dynamic compression plates (DCP) in goat tibiae. Methods: Twenty pairs of matched fresh goat tibiae were used. A transverse fracture model was established. The fractures with a 3mm interspace between the fracture ends were subject to fixations with the DCPs and the PC-DCPs respectively, then the four-points bending tests and the torsion tests were conducted to compare the mechanical properties of the PC-DCP with that of DCP. Another 13 sexually mature goats underwent fixations with the DCPs and the PC-DCPs, respectively, at the mid-shafts of the intact bilateral tibiae. Ischemic zones were observed at four time points (1 day ,2,6, and 12 weeks after operation) using disulphine blue staining technique. Results: There were no significant differences in mechanical properties, such as bend- and torsion-resistance, between the DCPs and the PC-DCPs. One day, 2, and 6 weeks after operation, on the side of DCP fixation, outer cortical bone ischemia under the plate persisted, and this condition did not reverse until 12 weeks after operation. However, on the side of PC-DCP fixation, cortical bone ischemia occurred only in the periphery of the screw holes and at the contact sites of the PC NUTs 1 day after operation, and it disappeared at 2 weeks after operation. Conclusions: The PC-DCP has similar biomechanical properties of the DCP, byt is less detrimental to local bone blood circulation than the conventional plates.
Keywords:Fracture fixation   internal  Bone plates  Microcirculation