Ultrasound Densitometric Analysis: Comparison Between an Online Digital Acquisition Acoustic Program and an Offline Analog Program |
| |
Authors: | Fernandez Anthony P. Aronson Solomon Toledano Alicia Winkelmann Jacqueline Saldivar Jose Feinstein Steven B. |
| |
Affiliation: | Department of Anesthesia and Critical Care, University of Chicago, and the Section of Cardiology, University of Illinois at Chicago, Chicago, Illinois |
| |
Abstract: | Videodensitometric analysis of myocardial contrast echocardiography is traditionally performed off line. Recently, an online contrast ultrasound analysis system, Acoustic Densitometry (Hewlett-Packard), was introduced. We compared pixel intensities acquired with Acoustic Densitometry to pixel intensities derived from videodensitometry. A tissue phantom was imaged in phase I using three transducer frequencies (2.5, 3.5, and 5.0 MHz). In phase II, an in vitro flowing tube model with various concentrations of Albunex® was imaged at two flow rates, 0.6 and 1.2 m/sec, and at two transducer frequencies, 2.5 and 3.5 MHz. The relationship between pixel intensities yielded by the two systems for identical ultrasound signals was determined with linear regression. Intensities derived with Acoustic Densitometry strongly correlated with those derived from the offline videodensitometry system. The intensities were related by a predictive multiplicative factor based on display characteristics of the two systems. These results suggest that semiquantitative, online perfusion analysis with Acoustic Densitometry is as sensitive as analysis offline with videodensitometry. |
| |
Keywords: | contrast echocardiography pixel intensity myocardium videodensitometry |