Abstract: | New Clambda-values (Cair,lambda) are proposed which should be applied for ionization chambers with an inner wall of air-equivalent material, air eq. plastic or graphite. The new values are up to approximately 3 per cent lower, and apply for a chamber with an inner wall lining of water-equivalent material.er cent higher than those published in the ICRU Report No. 14 and here named Cwater,lambda as the inner wall is considered water-equivalent. Also two sets of CE-values are proposed, namely Cair,E which is given in ICRU No. 21 and Cwater,E which is approximately 3 p |