首页 | 本学科首页   官方微博 | 高级检索  
     


Flucytosine interference with serum creatinine determinations
Authors:M L Santeiro  D F Thompson  R Sagraves
Affiliation:University of Oklahoma College of Pharmacy, Oklahoma City 73190.
Abstract:
A patient receiving intravenous amphotericin B and oral flucytosine was found to have falsely elevated serum creatinine values. Flucytosine has been reported to interfere with serum creatinine determinations when measured by the Kodak Ektachem-700 analyzer but not when the Jaffe reaction is employed. Serum creatinine values were determined by the two methods on six serum samples obtained from this patient at various times throughout her hospitalization. Flucytosine can cause clinically significant false elevations in serum creatinine when measured by the Kodak Ektachem-700 analyzer.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏