Development of an improved multi-pressure-sensor probe for recording muscle contraction in human intestine |
| |
Authors: | Dr. John R. Mathias MD Charles A. Sninsky MD Huntly D. Millar MSc Mary H. Clench PhD Richard H. Davis PA |
| |
Affiliation: | (1) Medical Service, Veterans Administration Medical Center, 32602 Gainesville, Florida;(2) the Department of Medicine, University of Florida, 32610 Gainesville, Florida;(3) Millar Instruments, Inc., 77023 Houston, Texas |
| |
Abstract: | We have developed an improved semiconductor recording probe for monitoring motility of the upper small intestine in humans. The probe consists of six ultraminiature silicon pressure sensors spaced 10 cm apart and encased in a flexible polyurethane sheath. The probe is small [2.67 mm (8F) diameter], is easily passed transnasally, and is tolerated by patients for prolonged recording periods (24–36 hr) with a minimum of discomfort. The initial semiconductor and catheter material were those designed for use in the cardiovascular system, but they proved to be easily damaged by gastric acid and enzymes. After improvement of this probe, we now have recordings from more than 100 patients for an approximate total of 6000 hr of recording time. The improved probe is a durable recording device that facilitates the investigation of motility of the small intestine in humans in health and disease.This study was supported by funds from the Medical Research Service of the Veterans Administration and by National Institutes of Health Clinical Research Center grant RR-82. |
| |