首页 | 本学科首页   官方微博 | 高级检索  
     


Contralateral loop snare removal of a ruptured and entrapped angioplasty balloon
Authors:Michael A. Braun M.D.  Steven J. Smith  Timothy N. Merrill
Affiliation:(1) Department of Radiology/Olson Pavilion, Northwestern Memorial Hospital, 710 North Fairbanks Court, Chicago, IL 60611, USA, US;(2) Department of Radiology, La Grange Memorial Hospital, 5105 Willow Springs Road, La Grange, IL 60525, USA, US
Abstract:
We describe a technique that extracted a ruptured angioplasty balloon which had become entrapped by a calcified left common iliac artery stenosis. The balloon catheter had been advanced crossover from the right and could not be retracted directly into a sheath across the aortic bifurcation. Therefore, a guidewire was inserted through the balloon catheter and captured by a loop snare advanced from the left femoral artery. The loop snare was also used to free the balloon wings from the stenosis. The balloon was then pulled into a 10 Fr sheath and removed as a unit with the sheath.
Keywords:: Arteries, iliac—  Transluminal angioplasty—  Stenosis or obstruction—  Balloon rupture—  Retrieval
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏