3-D real-time motion correction in high-intensity focused ultrasound therapy |
| |
Authors: | Pernot Mathieu Tanter Mickaël Fink Mathias |
| |
Affiliation: | Laboratoire Ondes et Acoustique, ESPCI, Université Paris VII, Paris, France. |
| |
Abstract: | A method for tracking the 3-D motion of tissues in real-time is combined with a 2-D high-intensity focused ultrasound (US), or HIFU, multichannel system to correct for respiratory motion during HIFU therapy. Motion estimation is based on an accurate ultrasonic speckle-tracking method. A pulse-echo sequence is performed for a subset of the transducers of the phased array. For each of these subapertures, the displacement is estimated by computing the 1-D cross-correlation of the backscattered signals acquired at two different times. The 3-D motion vector is then computed by a triangulation algorithm. This technique is experimentally validated in phantoms moving as fast as 40 mm s(-1), and combined with HIFU sequences. A real-time feedback correction of the HIFU beam is achieved by adjusting the delays of each channel. The sonications "locked on target" are interleaved with very short motion-estimation sequences. Finally, in vitro experiments of "locked on target" HIFU therapy are performed in fresh moving tissues. |
| |