首页 | 本学科首页   官方微博 | 高级检索  
     


Clinical oral malodor measurement with a portable sulfide monitor
Authors:Ueno M  Shinada K  Yanagisawa T  Mori C  Yokoyama S  Furukawa S  Takehara S  Kawaguchi Y
Affiliation:Department of Oral Health Promotion, Graduate School, Tokyo Medical and Dental University, Tokyo, Japan
Abstract:
Objective:  The aim of this study was to evaluate the clinical ability of Breathtron® by comparing it with other malodor measurement procedures: the organoleptic test (OT) and gas chromatography (GC).
Subjects and methods:  Patients were 475 patients who visited a fresh breath clinic. Oral malodor was measured with the OT, GC, and Breathtron®. Correlation analysis and two linear regression analyses were conducted to examine the relationship of the Breathtron® values with OT scores and volatile sulfide compound (VSC) concentrations by GC: i.e. the regression of Breathtron® on OT and the regression of Breathtron® on total VSCs by GC. Receiver operating characteristics (ROC) analysis was conducted to investigate the sensitivity and specificity of Breathtron®.
Results:  The Breathtron® values were significantly correlated with OT and VSCs by GC. In the regression analysis, predicted Breathtron® values were 199.3 and 520.1 ppb for OT scores 1 and 2, and reasonably close to total VSCs by GC between 550 and 750 ppb. The ROC analysis demonstrated that Breathtron® is a useful and valuable adjunct measurement tool.
Conclusions:  Breathtron® is a simple, rapid and reliable appliance for screening oral malodor if an appropriate malodor threshold level is chosen.
Keywords:oral malodor    organoleptic test    gas chromatography    Breathtron®
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号