The dosimetric properties of an intraoperative radiation therapy applicator system for a Mevatron-80 |
| |
Authors: | C E Nelson R Cook S Rakfal |
| |
Affiliation: | Department of Radiation Oncology, East Carolina University School of Medicine, Greenville, North Carolina 27858. |
| |
Abstract: | ![]() An applicator system for intraoperative radiation therapy has been fabricated which does not require physical docking with the accelerator. A dosimetric study has been completed which documents the properties of this system for a variety of electron beam energies, applicator sizes, collimator settings, both primary and secondary, and source-surface distance (SSD) settings. Sensitivity of the system to common misalignment errors was also determined. Results indicate (a) applicator leakage of less than 5%, (b) beam flatness to within plus or minus 5% at the dMAX with a single primary collimator setting, (c) smooth changes in output with cone size, beam energy and SSD, and (d) negligible changes in dose distributions within alignment errors permitted by the system. |
| |