首页 | 本学科首页   官方微博 | 高级检索  
     


MR-safety in clinical practice at 7T: Evaluation of a multistep screening process in 1819 subjects
Affiliation:1. Department of Medical Imaging and Physiology, Skåne University Hospital, Lund, Sweden;2. Department of Diagnostic Radiology, Clinical Sciences, Lund University, Lund, Sweden;3. Lund University Bioimaging Center (LBIC), Lund University, Lund, Sweden;4. Karolinska University Hospital, Solna Stockholm, Sweden;5. Department of Radiology, Institute of Clinical Sciences, Sahlgrenska Academy, University of Gothenburg and Sahlgrenska University Hospital, Gothenburg, Sweden
Abstract:
IntroductionMR facilities must implement and maintain adequate screening and safety procedures to ensure safety during MR examinations. The aim of this study was to evaluate a multi-step MR safety screening process used at a 7T facility regarding incidence of different types of safety risks detected during the safety procedure.MethodsSubjects scheduled for an MR examination and having entered the 7T facility during 2016–2019 underwent a pre-defined multi-step MR safety screening process. Screening documentation of 1819 included subjects was reviewed, and risks identified during the different screening steps were compiled. These data were also related to documented decisions made by a 7T MR safety committee and reported MR safety incidents.ResultsPassive or active implants (n = 315) were identified in a screening form and/or an additional documented interview in 305 subjects. Additional information not previously self-reported by the subject, regarding implants necessitating safety decisions performed by the staff was revealed in the documented interview in 102 subjects (106 items). In total, the 7T MR safety committee documented a decision in 36 (2%) of the included subjects. All of these subjects were finally cleared for scanning.ConclusionA multi-step screening process allows a thorough MR screening of subjects, avoiding safety incidents. Different steps in the process allow awareness to rise and items to be detected that were missed in earlier steps.Implications for practiceSafety questions posed at a single timepoint during an MR screening process might not reveal all safety risks. Repetition and rephrasing of screening questions leads to increased detection of safety risks. This could be effectively mitigated by a multi-step screening process. A multi-disciplinary safety committee is efficient at short notice responding to unexpected safety issues.
Keywords:Safety management  Magnetic resonance imaging  Patient safety  Health care  Documentation  Implants  ACR"  },{"  #name"  :"  keyword"  ,"  $"  :{"  id"  :"  kwrd0045"  },"  $$"  :[{"  #name"  :"  text"  ,"  _"  :"  American college of radiology  MR"  },{"  #name"  :"  keyword"  ,"  $"  :{"  id"  :"  kwrd0055"  },"  $$"  :[{"  #name"  :"  text"  ,"  _"  :"  Magnetic resonance  RIS"  },{"  #name"  :"  keyword"  ,"  $"  :{"  id"  :"  kwrd0065"  },"  $$"  :[{"  #name"  :"  text"  ,"  _"  :"  Radiological information system  SF"  },{"  #name"  :"  keyword"  ,"  $"  :{"  id"  :"  kwrd0075"  },"  $$"  :[{"  #name"  :"  text"  ,"  _"  :"  Screening form
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号