首页 | 本学科首页   官方微博 | 高级检索  
     


Systematic evaluation of three different commercial software solutions for automatic segmentation for adaptive therapy in head-and-neck, prostate and pleural cancer
Authors:Mariangela La Macchia  Francesco Fellin  Maurizio Amichetti  Marco Cianchetti  Stefano Gianolini  Vitali Paola  Antony J Lomax  Lamberto Widesott
Affiliation:1. Agenzia Provinciale per la Protonterapia, Via F.lli Perini, 181, 38122, Trento, Italy
3. Center for Proton Radiation Therapy, Paul Scherrer Institute, Villigen PSI, Switzerland
2. Istituto del Radio ??O. Alberti??, Spedali Civili, Brescia, Italy
4. Department of Physics, Swiss Institute of Technology (ETH), Zurich, Switzerland
Abstract:

Purpose

To validate, in the context of adaptive radiotherapy, three commercial software solutions for atlas-based segmentation.

Methods and materials

Fifteen patients, five for each group, with cancer of the Head&Neck, pleura, and prostate were enrolled in the study. In addition to the treatment planning CT (pCT) images, one replanning CT (rCT) image set was acquired for each patient during the RT course. Three experienced physicians outlined on the pCT and rCT all the volumes of interest (VOIs). We used three software solutions (VelocityAI 2.6.2 (V), MIM 5.1.1 (M) by MIMVista and ABAS 2.0 (A) by CMS-Elekta) to generate the automatic contouring on the repeated CT. All the VOIs obtained with automatic contouring (AC) were successively corrected manually. We recorded the time needed for: 1) ex novo ROIs definition on rCT; 2) generation of AC by the three software solutions; 3) manual correction of AC. To compare the quality of the volumes obtained automatically by the software and manually corrected with those drawn from scratch on rCT, we used the following indexes: overlap coefficient (DICE), sensitivity, inclusiveness index, difference in volume, and displacement differences on three axes (x, y, z) from the isocenter.

Results

The time saved by the three software solutions for all the sites, compared to the manual contouring from scratch, is statistically significant and similar for all the three software solutions. The time saved for each site are as follows: about an hour for Head&Neck, about 40?minutes for prostate, and about 20?minutes for mesothelioma. The best DICE similarity coefficient index was obtained with the manual correction for: A (contours for prostate), A and M (contours for H&N), and M (contours for mesothelioma).

Conclusions

From a clinical point of view, the automated contouring workflow was shown to be significantly shorter than the manual contouring process, even though manual correction of the VOIs is always needed.
Keywords:
本文献已被 SpringerLink 等数据库收录!