Analysis of glistenings in hydrophobic acrylic intraocular lenses on visual performance |
| |
Authors: | Lei Xi Yi Liu Feng Zhao Chuan Chen Bing Cheng |
| |
Affiliation: | State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China;State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China;State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China;State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China;State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou 510060, Guangdong Province, China |
| |
Abstract: |
AIMTo assess patients'' visual performance with glistenings in one piece soft hydrophobic acrylic intraocular lenses (IOLs) (Alcon) 2 years postoperatively.METHODSThis cross section trial included 120 eyes with one piece IOL at 2 years postoperatively. Glistening was classified in 4 groups, ranging from 0 (none) to 3 (most evident) according to their severity in IOLs optics observed under a slit lamp. All eyes underwent a uncorrected and best-corrected visual acuity evaluation (UCVA and BCVA, LogMAR scale), a complete clinical examination, a contrast sensitivity (CS) evaluation by F.A.C.T chart, and a visual field test by Humphrey Field Analyzer ‖ (HFA). One-way ANOVA was used for quantitative data, while Pearson χ2 test was used for qualitative data to analyze the visual function of 4 glistening groups.RESULTSTotally 120 eyes were enrolled with 30 eyes in each glistening group. There was no statistical correlation between glistening grades and patients'' age, IOLs power, postoperative UCVA and BCVA (P>0.05). Quantificationally, CS values among each group were not statistically different. However, qualitative analysis showed there were more eyes in grade 3 group than in grade 0 group having abnormally declined CS at high spatial frequency (10% vs 36.7% at 18 cpd, P=0.029; 6.7% vs 26.7% at 12 cpd, P=0.013). Mean deviation (MD) of the visual field test was -2.14±2.31, -1.97±2.23, -3.02±3.17, -4.12±3.38 in group 0 to 3 respectively. There was a significant decrease in the most serious glistenings group (P =0.018).CONCLUSIONGlistenings may potentially have an impact on contrast sensitivity at high spatial frequency and MD in visual field test. |
| |
Keywords: | cataract intraocular lens hydrophobic acrylic contrast sensitivity mean deviation |
|
| 点击此处可从《国际眼科》浏览原始摘要信息 |
|
点击此处可从《国际眼科》下载免费的PDF全文 |
|