Sensitivity of a Direct Computer-aided Detection System in Full-field Digital Mammography for Detection of Microcalcifications Not Associated with Mass or Architectural Distortion
Authors:Anabel M. Scaranelo  Pavel Crystal  Karina Bukhanov  Thomas H. Helbich
Affiliation:1. Department of Medical Imaging, Princess Margaret Hospital, University of Health Network and Mount Sinai Hospital, University of Toronto, Toronto, Ontario, Canada;2. Department of Radiology, Medical University of Vienna, Vienna, Austria
Abstract:

Purpose

The purpose of this study was to evaluate the sensitivity of a direct computer-aided detection (CAD) system (d-CAD) in full-field digital mammography (FFDM) for the detection of microcalcifications not associated with mass or architectural distortion.

Materials and Methods

A database search of 1063 consecutive stereotactic core biopsies performed between 2002 and 2005 identified 196 patients with Breast Imaging-Reporting and Data System (BI-RADS) 4 and 5 microcalcifications not associated with mass or distortion detected exclusively by bilateral FFDM. A commercially available CAD system (Second Look, version 7.2) was retrospectively applied to the craniocaudal and mediolateral oblique views in these patients (mean age, 59 years; range, 35–84 years). Breast density, location and mammographic size of the lesion, distribution, and tumour histology were recorded and analysed by using χ2, Fisher exact, or McNemar tests, when applicable.

Results

When using d-CAD, 71 of 74 malignant microcalcification cases (96%) and 101 of 122 benign microcalcifications (83%) were identified. There was a significant difference (P < .05) between CAD sensitivity on the craniocaudal view, 91% (68 of 75), vs CAD sensitivity on the mediolateral oblique view, 80% (60 of 75). The d-CAD sensitivity for dense breast tissue (American College of Radiology [ACR] density 3 and 4) was higher (97%) than d-CAD sensitivity (95%) for nondense tissue (ACR density 1 and 2), but the difference was not statically significant. All 28 malignant calcifications larger than 10 mm were detected by CAD, whereas the sensitivity for lesions small than or equal to 10 mm was 94%.

Conclusions

D-CAD had a high sensitivity in the depiction of asymptomatic breast cancers, which were seen as microcalcifications on FFDM screening, with a sensitivity of d-CAD on the craniocaudal view being significantly better. All malignant microcalcifications larger than 10 mm were detected by d-CAD.
Keywords:Breast cancer   Computer-aided detection   Computer-aided diagnosis   Digital images   Full-field digital mammography   Microcalcifications