Factors causing differences in voiding parameters between conventional and ambulatory urodynamics |
| |
Authors: | J. Groen R. van Mastrigt R. Bosch |
| |
Affiliation: | (1) Department of Urology – urodynamics, Room EE 1630, Erasmus University Rotterdam, P.O. Box 1738, 3000 DR Rotterdam, The Netherlands Tel.: +31-10-4087377; Fax: +31-10-4089451, NL |
| |
Abstract: | Voiding parameter values measured with ambulatory urodynamic monitoring (AM) are generally found to be different from those measured with conventional cystometry (CMG). The reason for this is unclear, but might be related to differences in the voided volume. To verify this hypothesis, we compared voidings from female patients at an initial bladder volume that was close to the modal volume (that is, the volume most often voided by the patient as derived from frequency/volume charts) with voidings at maximum cystometric capacity during a routine video urodynamic examination. A first group of 35 patients voided at the modal volume before they did at capacity. The order was reversed in a second group of 12 patients. The dependence of the voiding parameters on the voided volume and the order of the measurements were examined. It was found that the maximum flow rate depended significantly on the voided volume, but the associated detrusor pressure did not. Urethral resistance and bladder contraction strength were not volume dependent either. It was concluded that the differences between AM and CMG cannot be explained from possible differences in the voided volume. Received: 23 August 1999 / Accepted: 16 December 1999 |
| |
Keywords: | Urodynamics Cystometry Bladder volume Pressure/flow analysis Ambulatory monitoring |