Assessment of endoleak significance after endovascular repair of abdominal aortic aneurysms: a lumped parameter model |
| |
Authors: | Wolters B J B M Emmer M Rutten M C M Schurink G W H van de Vosse F N |
| |
Affiliation: | aDepartment of Biomedical Engineering, Eindhoven University of Technology, The Netherlands bDepartment of Surgery, University Hospital Maastricht, The Netherlands |
| |
Abstract: | The outcome of endovascular repair of abdominal aortic aneurysms (AAAs) is greatly compromised by the possible occurrence of endoleak. Previously, the causes and effects of endoleak on a patient-specific basis have mainly been investigated in experimental studies. In order to both reconcile and physically substantiate the various experimental findings, a lumped parameter model of an incompletely excluded AAA was developed. After experimental validation, the model was applied to study the effects on the intrasac pressure of the degree of endoleak, the degree of stent-graft compliance, and the resistance of a possible outflow tract formed by a branching vessel. It is concluded that the presence of endoleak leads to elevated intrasac pressure, the mean of which is mainly governed by the outflow tract resistance, while the pulse pressure is governed by both the endoleak resistance and the stent-graft compliance. Based on the agreement of the current results with previous findings, it is further concluded that the lumped parameter modelling method provides a useful numerical tool for validating experimental endoleak studies. |
| |
Keywords: | Abdominal aortic aneurysm Endoleak Intrasac pressure Experimental setup Computational model |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|