Kodak EDR2 film for patient skin dose assessment in cardiac catheterization procedures |
| |
Authors: | Morrell R E Rogers A T |
| |
Affiliation: | Medical Physics Directorate, Nottingham City Hospital NHS Trust, Hucknall Road, Nottingham NG5 1PB, UK. |
| |
Abstract: | Patient skin doses were measured using Kodak EDR2 film for 20 coronary angiography (CA) and 32 percutaneous transluminal coronary angioplasty (PTCA) procedures. For CA, all skin doses were well below 1 Gy. However, 23% of PTCA patients received skin doses of 1 Gy or more. Dose-area product (DAP) was also recorded and was found to be an inadequate indicator of maximum skin dose. Practical compliance with ICRP recommendations requires a robust method for skin dosimetry that is more accurate than DAP and is applicable over a wider dose range than EDR2 film. |
| |