首页 | 本学科首页   官方微博 | 高级检索  
     


Scintigraphic diagnosis of tricuspid regurgitation
Authors:Tu'meh, SS   Tracy, DA   Wynne, J   Konstam, MA   Kozlowski, JF   Neumann, AL   Holman, BL
Abstract:
The authors describe a simple technique for diagnosis of tricuspid regurgitation. Red blood cells were labeled in vivo with 99mTc and 22 patients were studied with ECG-gated blood-pool imaging of the liver. A single region of interest was manually drawn around the liver and a time-activity curve obtained. The per cent change in liver counts during the cardiac cycle was found to be significantly higher in the 12 patients with tricuspid regurgitation (Group I) (mean, 4.04 +/- 1.6%; range, 1.3-21.4%) compared with the 10 controls (Group II) (mean, 0.35 +/- 0.16%; range, 0.013-1.3%) (p less than 0.05). Using a 1% change in liver counts as the criterion of a positive study, all 12 cases in Group I were diagnosed correctly, but there was one false positive in Group II; thus the sensitivity was 100% and the specificity 90%.
Keywords:
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号