Electrocardiogram shows reliable heart rates much earlier than pulse oximetry during neonatal resuscitation |
| |
Authors: | Hiroshi Mizumoto Seiichi Tomotaki Hirofumi Shibata Kazutoshi Ueda Ryoko Akashi Hiroko Uchio Daisuke Hata |
| |
Affiliation: | Department of Pediatrics, Kitano Hospital, Tazuke Kofukai Medical Research Institute, Osaka, Japan |
| |
Abstract: | Background: The aim of this study was to determine the usefulness of the three‐lead electrocardiogram (ECG) during neonatal resuscitation. Methods: Both pulse oximetry (PO) and ECG were applied immediately after delivery to measure heart rate (HR). We reviewed video recordings of the respective monitors, and checked the time at which each monitor started to display reliable rate values. Results: In 20 deliveries, ECG showed HR much earlier than PO (median 38 s vs 122 s after delivery). ECG displayed reliable HR throughout resuscitation. We were able to confirm the effectiveness of the initial respiratory support from the elevation in HR. Conclusions: ECG was a safe and reliable method for showing HR, and was used to determine the initiation and the effectiveness of resuscitation in the delivery room. |
| |
Keywords: | electrocardiogram heart rate neonatal resuscitation pulse oximetry |
|
|