首页 | 本学科首页   官方微博 | 高级检索  
     


Use of pulse oximetry for monitoring tracheostomy tube obstruction
Authors:A J Rao  T Komatsu  M Kainuma  T Koh  Y Shimada  N Yanagita  T Nakashima
Affiliation:Department of Otorhinolaryngology, Nagoya University School of Medicine, Japan.
Abstract:
Experimental studies were performed on dogs for monitoring postoperative tracheostomy tube obstruction using a pulse oximeter. While a 25% obstruction of the cross-sectional area of the tube did not alter the oxygen saturation, obstruction of 40% and above resulted in a drop in the oxygen saturation level which corresponded to the degree of obstruction. The greater the degree of obstruction, the sooner was the fall of oxygen saturation. Therefore a pulse oximeter could be effectively put to use to monitor tracheostomy tube obstruction.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号