Icare回弹式眼压计与Goldmann压平眼压计眼压测量结果的比较
引用本文:周和政,张绍阳,江文珊.Icare回弹式眼压计与Goldmann压平眼压计眼压测量结果的比较[J].眼科研究,2011,29(5):456-459.
作者姓名:周和政  张绍阳  江文珊
作者单位:广州军区武汉总医院眼科,武汉,430070
摘    要:背景Icare回弹式眼压计作为一种新式眼压计,有必要对它的临床应用价值进行评估。目的通过比较分析Icare回弹式眼压计和Goldmann压平眼压计(GAT)的眼压测量结果,探讨Icare的临床价值。方法可疑青光眼、青光眼、屈光不正及部分健康体检者78例共152眼同时接受Icare、GAT眼压测量,受检眼先行Icare测量,然后再进行GAT测量,2次测量间隔3~5min。对比分析两种眼压计的测量结果。结果使用Icare和GAT测得的眼压均值分别为(19.16±5.03)mmHg和(18.41±4.52)mmHg,96眼(63.2%)两者的眼压差值≤1mmHg,二者的测量值差异虽有统计学意义,但二者的变化呈明显正相关(r=0.940,P〈0.01)。当Icare眼压测量值〈16mmHg时,Icare的眼压测量值低于GAT,而当Icare眼压测量值≥16mmHg时恰好相反;CCT偏薄、正常以及偏厚的情况下,Icare的眼压测量值均高于GAT的眼压测量值。Icare、GAT的眼压测量值和CCT间呈正相关(r=0.341,P〈0.01;r=0.333,P〈0.01)。结论与GAT眼压计比较,Icare回弹式眼压计易操作,测量结果可靠,临床实用性更强。

关 键 词:Icare回弹式眼压计  眼压  中央角膜厚度

Comparison of intraocular pressure measured by Icare rebound tonometer and Goldmann applanation tonometer
ZHOU He-zheng,ZHANG Shao-yang,JIANG Wen-shan.Comparison of intraocular pressure measured by Icare rebound tonometer and Goldmann applanation tonometer[J].Chinese Ophthalmic Research,2011,29(5):456-459.
Authors:ZHOU He-zheng  ZHANG Shao-yang  JIANG Wen-shan
Institution:.( Department of Ophthalmology, Wuhan General Hospital of Chinese PLA , Guangzhou Military Command, Wuhan 430070, China)
Abstract:Background As a new tonometer,it is necessary to assess the clinical value of Icare rebound tonometer.Objective This study was to compare the intraocular pressure(IOP)values measured by Icare with that measured by GAT,and discuss the clinical value of leare rebound tonometer. Methods IOP measurement was performed on 152 eyes of 78 subjects with suspicious glaucoma,glaucoma,refractive error and normal examinnee by Icare and GAT respectively.The Icare IOP was measured firstly and then the GAT IOP was carried out with the 3-or 5-minute interval.The IOP values were compared between Ieare and GAT.This study was approved by Ethic Committee of Wuhan General Hospital of Chinese PLA.Written informed consent was obtained from each subject prior to this study. Results The mean IOP values of Icare and GAT were(19.16±5.03)mmHg and(18.41±4.52)mmHg respectively.The differences between Icare IOP and CAT IOP were less than or equal to 1 mmHg in 96 of 105 eyes(63.2%).The positive correlation was found between the Icare IOP and GAT IOP(r=0.940,P<0.01).The Ieare IOP was lower than that of GAT when IOPIcare<16 mmHg,however,the IOP of Icare were higher when IOPIcare≥6 mmHg;the IOP of Icare were higher than that of GAT in the total CCT range.The correlation coefficients of IOP of Icare or CAT with CCT were 0.341(P<0.01)and 0.333(P<0.01),respectively. Conclusion Compared with GAT,Icare is more feasible in clinic because it is practicable and reliable.
Keywords:Icare rebound tonometer  Intraocular pressure  Central corneal thickniss