首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of Vancomycin Susceptibility Testing for Methicillin-Resistant Staphylococcus aureus: Comparison of Etest and Three Automated Testing Methods
Authors:Michael J. Rybak  Celine Vidaillac  Helio S. Sader  Paul R. Rhomberg  Hossein Salimnia  Lawrence E. Briski  Audrey Wanger  Ronald N. Jones
Abstract:We evaluated the ability of four commercial MIC testing systems (MicroScan, Vitek 2, Phoenix, and Etest) to detect vancomycin MIC values of ≤1 to ≥2 in 200 methicillin-resistant Staphylococcus aureus (MRSA) strains compared to the Clinical and Laboratory Standards Institute broth microdilution (BMD) reference methods. Compared to the BMD method, absolute agreement (0 ± dilution) was highest for the Phoenix system (66.2%) and the MicroScan turbidity method (61.8%), followed by the Vitek 2 system (54.3%). The Etest produced MIC values 1 to 2 dilutions higher than those produced by the BMD method (36.7% agreement). Of interest, the MicroScan system (prompt method) was more likely to overcall an MIC value of 1 mg/liter (74.1%), whereas the Phoenix (76%) and Vitek 2 (20%) systems had a tendency to undercall an MIC of 2 mg/liter. The ability to correctly identify vancomycin MIC values of 1 and 2 has clinical implications and requires further evaluation.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号