MOSFET dosimetry for radiation dose assessment of bismuth shielding of the eye in children |
| |
Authors: | Mukundan Srinivasan Wang Page Inman Frush Donald P Yoshizumi Terry Marcus Jeffrey Kloeblen Emily Moore Meredith |
| |
Affiliation: | Department of Radiology, Duke University Medical Center, DUMC Box 3808, Durham, NC 27705, USA. |
| |
Abstract: | OBJECTIVE: The purpose of our study was to measure radiation dose to the orbit during pediatric cranial CT with and without bismuth shielding using a novel dosimetry system. Cranial CT was performed on a pediatric anthropomorphic phantom, with and without bismuth eye shields. A solid-state metal oxide semiconductor field effect transistor (MOSFET) dosimeter was used to obtain real-time dose measurements. CONCLUSION: Bismuth shielding reduced radiation dose to the eye by up to 42%; shield artifact fell outside the diagnostic area of interest. |
| |