BC-6600型全自动血细胞分析仪性能评价
引用本文:李春碧,伏攀. BC-6600型全自动血细胞分析仪性能评价[J]. 检验医学与临床, 2014, 0(17): 2347-2349
作者姓名:李春碧  伏攀
作者单位:四川省广元市中心医院检验科 ,628000
基金项目:国家高技术研究发展计划(863计划)资助项目(2011AA02A111).
摘    要:
目的:对BC-6600型全自动血细胞分析仪(简称BC-6600分析仪)进行性能评价。方法对BC-6600型分析仪精密度、准确度、携带污染率、线性范围、可比性、白细胞分类计数等性能进行评价。结果 BC-6600型分析仪白细胞计数、红细胞计数、血红蛋白、血小板计数、网织红细胞计数结果具有良好的精密度和准确度,携带污染率分别为0.38%、0.48%、0.00%、0.98%、0.72%,在允许范围内;线性范围基本能覆盖临床标本浓度范围;与BC-5800型全自动血细胞分析仪检测结果具有良好的相关性;与手工分类计数相比,中性粒细胞、淋巴细胞和嗜酸粒细胞计数结果相关性良好,单核细胞计数结果相关性一般,嗜碱粒细胞计数结果相关性较差。结论 BC-6600型分析仪分析性能良好,能满足临床血液常规检测要求。

关 键 词:血细胞分析仪  精密度  携带污染率  线性范围

Performance evaluation of BC-6600 automated hematology analyzer
LI Chun-bi,FU Pan. Performance evaluation of BC-6600 automated hematology analyzer[J]. Laboratory Medicine and Clinic, 2014, 0(17): 2347-2349
Authors:LI Chun-bi  FU Pan
Affiliation:( Clinical Laboratory, Center Hospital of Ouangyuan, C-uangyuan, Sichuan 628000, China)
Abstract:
Objective To evaluate the performance of BC-6600 automated hematology analyzer .Methods Per-formance of precision ,accuracy ,carryover rate ,linearity ,comparability and differential counting were analyzed . Results The detection of white blood cells ,red blood cells ,hemoglobin ,platelet and reticulocytes were with fine pre-cision and accuracy ,and the carryover rates were 0 .38% ,0 .48% ,0 .00% ,0 .98% and 0 .72% .The linearity range covered the concentration range of clinical specimens .The detected results of BC-6600 and BC-5800 hematology ana-lyzer were with fine correlation .Compared with manual methods ,the counting of neutrophile granulocyte ,lymphocyte and eosnophils were fine correlated ,but the counting of monocyte was with low correlation .Conclusion The analysis performance of BC-6600 hematology analyzer could be fine ,which might be able to meet requirement for blood routine tests .
Keywords:hematology analyzer  precision  carryover rate  linearity
本文献已被 维普 等数据库收录!