Implantation of Active Can Implantable Defibrillators in the right Pectoral Region |
| |
Authors: | STEEN M. JENSEN ADRIAN PIETERSEN XU CHEN |
| |
Affiliation: | Cardiovascular Laboratory, Department of Medicine B, Rigshospitalet, University of Copenhagen, Copenhagen, Denmark |
| |
Abstract: | Routinely the active can ICD is placed in the left side pectoral position, which theoretically gives optimal conditions for a low defibrillation threshold. Some patients, bowever, demand a right pectoral position, which possibly could result in a bigger defibrillation threshold. A right pectoral position was used in 3 of 85 active can ICDs implanted in our institution from 1994. the DFT was 12 J in two and 18 f in one patient. Thus, right pectoral implantation is feasible and offers an alternative approach in selected patients. |
| |
Keywords: | ICD implantable cardioverter defibrillator active can DFT defibrillation threshold right pectoral position |
|
|