CentriMag left ventricular assist system: cannulation through a right minithoracotomy |
| |
Authors: | Gregoric Igor D Cohn William E Akay Mehmet H La Francesca Saverio Myers Tim Frazier O H |
| |
Affiliation: | Department of Cardiovascular Surgery, Texas Heart Institute at St. Luke's Episcopal Hospital, Houston, Texas 77030, USA. igregoric@heart.thi.tmc.edu |
| |
Abstract: | ![]() The CentriMag left ventricular assist system can be used for perioperative or postcardiotomy circulatory support of the failing heart. The device resides at the patient's bedside, and the cannulae are usually inserted through a midline sternotomy, with the inflow cannula in the left ventricle or right superior pulmonary vein and the outflow cannula in the aorta. In a patient whose chest has been closed and who has a delayed need for temporary mechanical support, a less invasive method of left ventricular assist device cannula insertion is preferred. In these cases, the CentriMag cannulae can be inserted through a right minithoracotomy with the inflow cannula in the right superior pulmonary vein and the outflow cannula in the aorta, with no heparinization. Herein, we describe this approach in a patient who experienced postcardiotomy cardiogenic shock after aortocoronary bypass surgery. This technique may facilitate ambulation and recovery in selected patients. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|