Delivery sheath tear after modification for ASD closure.
Authors:Mark H Hoyer
Affiliation:Section of Pediatric Cardiology, Indiana University School of Medicine, Indianapolis, 46202, USA. mhoyer@iupui.edu
Abstract:
During transcatheter closure of an atrial septal defect with insufficient aortic rim, a standard delivery sheath was modified by cutting a bevel at the distal tip to improve device orientation. The sheath split longitudinally when attempting to recapture the closure device. Troubleshooting allowed a device to be implanted successfully. Pitfalls regarding our sheath modification and methods to overcome prolapse of the left atrial disk into the right atrium are discussed.
Keywords:transcatheter ASD closure  delivery sheath modification