In Vivo Corrosion of Sleeved Ceramic Femoral Heads: A Retrieval Study |
| |
Authors: | Cody C Wyles Joshua M Kolz Douglas W Van Citters Daniel J Berry Robert T Trousdale |
| |
Institution: | 1. Department of Orthopedic Surgery, Mayo Clinic, Rochester, MN;2. Thayer School of Engineering, Dartmouth College, Hanover NH |
| |
Abstract: | BackgroundThe purpose of this study was to evaluate a series of retrieved sleeved ceramic femoral heads used in total hip arthroplasty (THA) and determine qualitative and quantitative damage and corrosion patterns.MethodsAn IRB-approved implant retrieval database was utilized to identify all sleeved ceramic femoral heads collected from 1995 to 2004. There were 16 implants with an average duration of in situ of 70 months (range, 13-241 months). The femoral stem was known in 14 cases and was titanium alloy in each of those cases. None were revised for metal-related complications. Ten implants (63%) were from primary THAs, and 6 (38%) were from revision THAs. Damage and corrosion were qualitatively graded using a modified Goldberg method. A quantitative assessment was performed with a coordinate measurement machine (CMM).ResultsAmong the 16 retrieved implants, 1 (6%) demonstrated severe Grade 4 corrosion, 5 (31%) had moderate Grade 3 corrosion, 5 (31%) had mild Grade 2 corrosion, and 5 (31%) had no visible corrosion at the inner sleeve that interfaces with the stem trunnion. The only case of grade 4 corrosion occurred in the only head-sleeve in the study that was not factory assembled and was mated with a titanium molybdenum zirconium ferrous (TMZF) alloy stem. The mean maximum linear corrosion depth at the taper interface, as measured by the CMM, was 7.7 microns (range, 0.9-32.9 microns).ConclusionThis study is the first to quantify corrosion at the titanium interface of sleeved ceramic femoral heads. Potentially clinically significant damage and corrosion patterns were observed in a few failed retrievals; however, the majority of cases demonstrated minimal or no damage. |
| |
Keywords: | total hip arthroplasty sleeved ceramic ceramic femoral head corrosion wear retrieval |
本文献已被 ScienceDirect 等数据库收录! |
|