首页 | 本学科首页   官方微博 | 高级检索  
     


Effect of transvenous cardiac resynchronization therapy device implantation on cardiac troponin I release
Authors:Altin Timucin  Akyurek Omer  Vurgun Kutay  Beton Osman  Sayin Tamer  Kilickap Mustafa  Karaoguz Remzi  Guldal Muharrem  Erol Cetin
Affiliation:Ankara University School of Medicine, Department of Cardiology, Ankara, Turkey. alitimaltin@yahoo.com
Abstract:
BACKGROUND: Pacemaker and implantable cardioverter defibrillator (ICD) implantation increases cardiac troponin I (cTnI) levels which indicates myocardial injury. During implantation of a cardiac resynchronization therapy (CRT) device, balloon inflation for coronary sinus (CS) venogram, cannulation of CS side branch, and electrode advancement may interfere with CS drainage and, hence, may decrease the washout of toxic metabolites from the heart. Thus, CRT implantation may further increase cTnI levels. In this study, we investigated the effects of CRT implantation on cTnI release. METHODS: We included 10 patients (mean age = 57 +/- 15 years) in whom a successful transvenous CRT system was implanted (CRT group). Twenty patients (mean age = 65 +/- 10 years) who underwent a transvenous pacemaker or ICD implantation were included as the control group. Blood samples for cTnI were drawn at baseline and at six, 12, 18, and 24 hours thereafter. RESULTS: Baseline median cTnI levels were similar in CRT and control groups (0.03 ng/mL vs 0.02 ng/mL, respectively; P = 0.1). Postoperative cTnI levels during 24 hours were significantly higher in the CRT group (P < 0.05) by two-way repeated measures of analysis of variance. Post hoc analysis revealed that cTnI levels were higher at the 6th, 12th, 18th, and 24th hours compared to baseline levels (P < 0.001, P < 0.001, P < 0.01, and P < 0.01, respectively). There was a significant difference in the area under the curves (AUCs) of cTnI measurements (1.79 hr.ng/mL in the CRT group and 0.78 hr.ng/mL in the control group, P < 0.05). CONCLUSION: Postoperative cTnI levels were higher after CRT implantation than simple pacemaker/ICD implantation. This may be due to CS manipulation during CRT implantation.
Keywords:cardiac resynchronization therapy    cardiac troponin I
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号