首页 | 本学科首页   官方微博 | 高级检索  
     


Development of silicone rubber hollow fiber membrane oxygenator for ECMO
Authors:Motomura Tadashi  Maeda Tomohiro  Kawahito Shiniji  Matsui Takahiro  Ichikawa Seiji  Ishitoya Hiroshi  Kawamura Masaki  Shinohara Toshiyuki  Sato Koshiro  Kawaguchi Yoichiro  Taylor Deborah  Oestmann Daniel  Glueck Julie  Nosé Yukihiko
Affiliation:Baylor College of Medicine, Michael E. DeBakey Department of Surgery, Houston, TX 77054, USA. Motomura@bcm.tmc.edu
Abstract:
Silicone rubber hollow fiber membrane produces an ideal gas exchange for long-term ECMO due to nonporous characteristics. The extracapillary type silicone rubber ECMO oxygenator having an ultrathin hollow fiber membrane was developed for pediatric application. The test modules were compared to conventional silicone coil-type ECMO modules. In vitro experiments demonstrated a higher O2 and CO2 transfer rate, lower blood flow resistance, and less hemolysis than the conventional silicone coil-type modules. This oxygenator was combined with the Gyro C1E3 centrifugal pump, and three ex vivo experiments were conducted to simulate pediatric V-A ECMO condition. Four day and 6 day experiments were conducted in cases 1 and 2, respectively. Case 3 was a long-term experiment up to 2 weeks. No plasma leakage and stable gas performances were achieved. The plasma free hemoglobin was maintained within a normal range. This compact pump-oxygenator system in conjunction with the Gyro C1E3 centrifugal pump has potential for a hybrid total ECMO system.
Keywords:Silicone rubber hollow fiber    ECMO    Plasma leakage    Centrifugal pump
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号