Analysis of Electromagnetic Exposure in Wireless Power Transfer System for Endoscopic Micro-robot |
| |
Authors: | YE Dong-dong YAN Guo-zheng LI Jia-jun WANG Kun-dong MA Guan-ying |
| |
Affiliation: | Institute of Precision Engineering and Intelligence Micro-system, School of Electronics and Information Technology, Shanghai Jiaotong University, Shanghai 200240, China |
| |
Abstract: | A wireless power transfer system for endoscopic micro-robot operating at 36 kHz is presented in this paper. The issue of patient's health and safety regarding exposure to the electromagnetic field is addressed. The specific absorption rate and current density can be used to investigate the electromagnetic influences on the biological tissues surrounded by the wireless power launching coil. In view of this purpose, the limited close-wound solenoid electromagnetic model is built, the relationship between the electric intensity and the specific absorption rate and current density is deduced,and the simulation experiments are done. Experimental results show that the values of SAR and current density related to different tissue catalogs are all very small and do not exceed their own limits respectively when the resonance frequency of operation is 36 kHz. |
| |
Keywords: | wireless power transfer endoscopic micro-robot specific absorption rate current density solenoid |
本文献已被 维普 万方数据 等数据库收录! |