首页 | 本学科首页   官方微博 | 高级检索  
     


An evaluation of a new single‐use flexible bronchoscope with a large suction channel: reliability of bronchoalveolar lavage in ventilated piglets and initial clinical experience
Authors:J. Mankikian  S. Ehrmann  L. Guilleminault  T. Le Fol  C. Barc  M. Ferrandière  T. Boulain  P. F. Dequin  A. Guillon
Affiliation:1. CHRU de Tours, Service de Réanimation Polyvalente, , Tours, France;2. INSERM, Centre d'Etude des Pathologies Respiratoires, UMR 1100/EA6305, , Tours, France;3. CHRU de Tours, Service de Pneumologie, , Tours, France;4. CHRU de Tours, Direction des services techniques et des investissements immobiliers, , Tours, France;5. INRA, UE1277, Plate‐Forme d'infectiologie Expérimentale, , Nouzilly, France;6. CHRU de Tours, Département d'Anesthésie‐Réanimation, , Tours, France;7. CHR Orléans, Service de Réanimation Médicale, , Orléans, France
Abstract:
A single‐use flexible bronchoscope with a large suction channel has become available recently and we have evaluated this innovative device. Firstly, bronchoalveolar lavage was performed and quantified in ventilated piglets. Next, the bronchoscope was evaluated in three intensive care units and a satisfaction questionnaire was carried out. Sixteen bronchoalveolar lavages were performed in piglets with a recovery rate of 83 (79–86 [72–89])% of the instilled volume. Quality and performance of all devices tested was identical. The medical satisfaction questionnaire was as follows: ‘acceptable’ to ‘very good’ for quality of aspiration, manoeuvrability and quality of vision; ‘very good’ to ‘perfect’ for setting up and insertion. This encouraging preliminary evaluation demonstrates the effectiveness of this new single‐use device, which may obviate the need for disinfection procedures and, thereby, eradicate a potential vector of patient cross‐contamination.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号