首页 | 本学科首页   官方微博 | 高级检索  
     


Relationship of fasting and hourly blood glucose levels to HbA1c values: safety, accuracy, and improvements in glucose profiles obtained using a 7-day continuous glucose sensor
Authors:Garg Satish  Jovanovic Lois
Affiliation:Barbara Davis Center for Childhood Diabetes, University of Colorado at Denver and Health Sciences Center, 1775 N. Ursula St. A140, Aurora, CO 80010, USA. satish.garg@uchsc.edu
Abstract:
OBJECTIVE: In this study, we evaluated the safety and efficacy of 7-day transcutaneous, real-time, continuous glucose monitoring (CGM) in subjects with insulin-requiring diabetes. RESEARCH DESIGN AND METHODS: Eighty-six subjects were enrolled at five U.S. centers. Subjects wore a sensor inserted under the skin of the abdomen for 7 days during each of three consecutive periods. Data were blinded during period 1 and unblinded during periods 2 and 3. RESULTS: Of the 6,811 matched self-monitoring of blood glucose to sensor values prospectively analyzed, 97.2% fell in the Clarke error grid zones A and B, and median absolute relative difference was 11.4%. After unblinding, subjects reduced time spent at <55 mg/dl by 0.3 h/day, reduced time spent at >240 mg/dl by 1.5 h/day, and increased time in the target zone (81-140 mg/dl) by 1.4 h/day (P < 0.05 for all three comparisons). Improvements were seen in both types 1 and 2 diabetes and with use of both multiple daily injections and continuous subcutaneous insulin infusion. Modal day graphs were generated in six groups of subjects based on HbA1c (A1C) (10%). Mean glucose levels from midnight to 7:00 a.m. (fasting and dawn phenomenon periods) were only normal for subjects with A1C
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号