首页 | 本学科首页   官方微博 | 高级检索  
     

磁力驱动轴流血泵溶血实验和动物实验
引用本文:吴广辉,蔺嫦燕,侯晓彤,张锡文,朱来来,李冰一,王景. 磁力驱动轴流血泵溶血实验和动物实验[J]. 生物医学工程与临床, 2010, 14(1): 1-4. DOI: 10.3969/j.issn.1009-7090.2010.01.001
作者姓名:吴广辉  蔺嫦燕  侯晓彤  张锡文  朱来来  李冰一  王景
作者单位:1. 首都医科大学附属北京安贞医院,北京心肺血管疾病研究所,北京,100029
2. 清华大学航天航空学院,北京,100084
基金项目:北京市青年科研资助项目,国家高技术研究发展计划(863计划) 
摘    要:目的对研制开发的一种新型的磁力外驱动轴流式心室辅助血泵的血液相容性能进行测试。方法利用特制血袋作为模拟循环管道,羊血作为循环介质,采用标准溶血指数衡量体外溶血实验性能。通过3只山羊12h在体实验衡量其在体适应性。结果实验测得轴流血泵体外实验标准溶血指数(NIH)为(0.158±0.043)mg/L。3例实验动物12h在体辅助无机械故障,血泵辅助后实验动物血液中游离血红蛋白(FHb)开始上升,最高达到164.8mg/L。结论磁力外驱动轴流血泵实验结果比较理想,值得进一步改进。

关 键 词:轴流血泵  溶血实验  动物实验

Hemolysis and animal experiments with the magnetically drived axial flow ventricular assist device
WU Guang-hui,LIN Chang-yan,HOU Xiao-tong,ZHANG Xi-wen,ZHU Lai-lai,LI Bing-yi,WANG Jing. Hemolysis and animal experiments with the magnetically drived axial flow ventricular assist device[J]. Biomedical Engineering and Clinical Medicine, 2010, 14(1): 1-4. DOI: 10.3969/j.issn.1009-7090.2010.01.001
Authors:WU Guang-hui  LIN Chang-yan  HOU Xiao-tong  ZHANG Xi-wen  ZHU Lai-lai  LI Bing-yi  WANG Jing
Affiliation:WU Guang-hui1,LIN Chang-yan1,HOU Xiao-tong1,ZHANG Xi-wen2,ZHU Lai-lai2,LI Bing-yi1,WANG Jing1 (1.Beijing Anzhen Hospital,Capital University of Medical Sciences,Beijing Institute of Cardiopulmonary , Vascular Disease,Beijing 100029,China,2. School of Aerospace,Tsinghua University,Beijing 100084,China)
Abstract:Objective To design and develop a magnetically drived axial left ventricular assist device,and evaluate its performance. Methods For the test of hemolysis in vitro mock circuit,a particular blood bag was used as a mock circulation loop and the sheep blood were used as the circulating medium. The normalized index of hemoglobin (NIH) was used to measure the hemolysis. The in vivo adaptability was tested through 12 hours in animal experiments with 3 goats. Results The NIH for the axial blood pump was (0.158 ± 0.043) mg/L. There were no mechanical breakdowns in 12 hours animal experiments; the free hemoglobin (FHb) of each experimental animal started rising just after circulation; the highest value was 164.8 mg/L. Conclusion The results obtained from the in vivo testing and animal experiment indicated that this is an acceptable design for blood pump,and worth to further improvement.
Keywords:axial blood pump  hemolysis testing  animal testing  
本文献已被 CNKI 维普 万方数据 等数据库收录!