首页 | 本学科首页   官方微博 | 高级检索  
     


Comparison of quantitative evaluation between cutaneous and transosseous inertial sensors in anterior cruciate ligament deficient knee: A cadaveric study
Authors:Atsunori Murase  Masahiro Nozaki  Masaaki Kobayashi  Hideyuki Goto  Masahito Yoshida  Sanshiro Yasuma  Tetsuya Takenaga  Yuko Nagaya  Jun Mizutani  Hideki Okamoto  Hirotaka Iguchi  Takanobu Otsuka
Affiliation:1. Department of Orthopaedic Surgery, Nagoya City University Graduate School of Medical Sciences, 1-Kawasumi, Mizuho-Cho, Mizuho-Ku, Nagoya 467-8601, Japan;2. Department of Joint Surgery for Rheumatic Disease, Nagoya City University Graduate School of Medical Sciences, 1-Kawasumi, Mizuho-Cho, Mizuho-Ku, Nagoya 467-8601, Japan;3. Department of Arthroplastic Medicine, Nagoya City University Graduate School of Medical Sciences, 1-Kawasumi, Mizuho-Cho, Mizuho-Ku, Nagoya 467-8601, Japan
Abstract:

Background

Recently several authors have reported on the quantitative evaluation of the pivot-shift test using cutaneous fixation of inertial sensors. Before utilizing this sensor for clinical studies, it is necessary to evaluate the accuracy of cutaneous sensor in assessing rotational knee instability. To evaluate the accuracy of inertial sensors, we compared cutaneous and transosseous sensors in the quantitative assessment of rotational knee instability in a cadaveric setting, in order to demonstrate their clinical applicability.

Methods

Eight freshly frozen human cadaveric knees were used in this study. Inertial sensors were fixed on the tibial tuberosity and directly fixed to the distal tibia bone. A single examiner performed the pivot shift test from flexion to extension on the intact knees and ACL deficient knees. The peak overall magnitude of acceleration and the maximum rotational angular velocity in the tibial superoinferior axis was repeatedly measured with the inertial sensor during the pivot shift test. Correlations between cutaneous and transosseous inertial sensors were evaluated, as well as statistical analysis for differences between ACL intact and ACL deficient knees.

Results

Acceleration and angular velocity measured with the cutaneous sensor demonstrated a strong positive correlation with the transosseous sensor (r = 0.86 and r = 0.83). Comparison between cutaneous and transosseous sensor indicated significant difference for the peak overall magnitude of acceleration (cutaneous: 10.3 ± 5.2 m/s2, transosseous: 14.3 ± 7.6 m/s2, P < 0.01) and for the maximum internal rotation angular velocity (cutaneous: 189.5 ± 99.6 deg/s, transosseous: 225.1 ± 103.3 deg/s, P < 0.05), but no significant difference for the maximum external rotation angular velocity (cutaneous: 176.1 ± 87.3 deg/s, transosseous: 195.9 ± 106.2 deg/s, N.S).

Conclusions

There is a positive correlation between cutaneous and transosseous inertial sensors. Therefore, this study indicated that the cutaneous inertial sensors could be used clinically for quantifying rotational knee instability, irrespective of the location of utilization.
Keywords:Corresponding author. Fax: +81 528420266.
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号