热固耦合式超声骨强度仪的可重复性及相关性研究 |
| |
引用本文: | 王 杰,倪朝民,陈焱焱,马祖长,孙怡宁,金 艳,张勤良. 热固耦合式超声骨强度仪的可重复性及相关性研究[J]. 中国康复医学杂志, 2012, 27(2): 125-129. DOI: 10.3969/i.issn.1001-1242.2012.02.007 |
| |
作者姓名: | 王 杰 倪朝民 陈焱焱 马祖长 孙怡宁 金 艳 张勤良 |
| |
作者单位: | 1. 安徽医科大学附属省立医院康复医学科,合肥,230000 2. 中国科学院合肥智能机械研究所 |
| |
基金项目: | 中国科学院知识创新工程领域前言重点项目(0923A11291) |
| |
摘 要: | ![]() 目的:评估一款国产热固耦合式超声骨强度仪的可重复性和有效性。方法:选择11名志愿者,用该仪器测得右跟骨的两个超声参数,分别是超声速度(SOS)和宽带超声衰减(BUA),计算每个参数的变异系数均方根百分比(CVRMS%),评估该仪器的短期可重复性;选择18名志愿者,用类似的方法评估该仪器的中期可重复性;选择29名志愿者,用双能X线吸收法(DEXA)测量腰椎骨密度(BMD),并与用该仪器测得的跟骨SOS,BUA值进行相关性分析,评估该仪器的有效性。结果:①超声参数SOS,BUA的短期CVRMS%分别为0.30,4.24;中期CVRMS%分别为0.41,4.20。②SOS与腰椎BMD的相关系数r=0.486(P<0.01),BUA与腰椎BMD的相关系数r=0.629(P<0.01)。结论:与国际主流定量超声(QUS)检测仪器相比,热固耦合式超声骨强度仪具有较高的可重复性和准确性,可用于临床骨质状况评估。
|
关 键 词: | 骨强度 定量超声 可重复性 相关性 双能X线吸收法 超声速度 宽带超声衰减 |
收稿时间: | 2011-05-15 |
Study on the repeatability and correlation of warm-air and solid coupled mode ultrasound bone strength device |
| |
Affiliation: | Affiliated Provincial Hospital of Anhui Medical University, Hefei, 230000 |
| |
Abstract: | ![]() Objective:To evaluate the repeatability and correlation of warm-air and solid coupled mode ultrasound bone strength device made in China.Method:Eleven and eighteen volunteers’ right calcaneus ultrasound parameters,speed of sound(SOS)and broadband ultrasound attenuation(BUA),were detected with the device,percentages of root-mean square of coefficient of variation(CVRMS%)of ultrasound parameters were calculated respectively,and used to assess the short-and middle-term repeatability of the device;29 volunteers’ right calcaneus ultrasound parameters(SOS,BUA)were detected with the device and bone mineral density(BMD)of lumbar spine were detected with dual energy X-ray absorptiometry(DEXA).Result:①The short-term CVRMS% were 4.24% for BUA,and 0.30% for SOS;the middle-term CVRMS% were 4.20% for BUA,and 0.41% for SOS.②The correlations between BMD of lumbar spine were 0.629(P<0.01)for BUA and 0.486(P<0.01)for SOS.Conclusion:Compared to international top-grade quantitative ultrasound(QUS)testing device,the warm-air and solid coupled mode ultrasound bone strength device has high performance of repeatability and accuracy,so it can be used to assessing bone mass status. |
| |
Keywords: | bone strength quantitative ultrasound repeatability correlation dual energy X-ray absorptiometry speed of sound broadband ultrasound attenuation |