儿童胸部DR降低辐射剂量的可行性分析
引用本文:郭辉,吴海忠,张鹏,曾群力,周晓珊,贺晓晔,李白艳. 儿童胸部DR降低辐射剂量的可行性分析[J]. 新疆医科大学学报, 2011, 34(11): 1254-1257
作者姓名:郭辉  吴海忠  张鹏  曾群力  周晓珊  贺晓晔  李白艳
作者单位:新疆医科大学第一附属医院放射科,乌鲁木齐,830011
基金项目:新疆医科大学第一附属医院专项基金
摘    要:
目的探讨儿童胸部数字化X线摄影(DR)降低辐射剂量的可行性。方法将0~14岁的儿童分为0~1岁、1~3岁、3~7岁、7~11岁、11~14岁5个年龄段,每个年龄段180例,共900例,随机分成A、B、C 3组,A组为Kodak DR 3000固定毫安秒组,B组为Kodak DR 3000自动毫安秒组,C组为Philips DR自动毫安秒组。对3组之间的毫安秒(mAs)均数(辐射剂量)和图像质量进行比较。结果 (1)3组mAs均值比较,C组比A组和B组分别降低4.89mAs、2.01mAs,降低幅度为81.5%、64.5%,3组毫安秒均值组间比较差异有统计学意义(F=16 814.679,P<0.01),即C组mAs均值最小,辐射剂量最低。(2)3组共900幅图像质量均能够满足临床诊断的要求,差异无统计学意义(χ2=5.095,P>0.05)。结论在获得满意图像质量前提下,Philips DR自动毫安秒组更能体现儿童摄影正当化及最优化原则。

关 键 词:辐射剂量  图像质量  儿童胸部  数字化X线摄影

Exploration of the feasibility on reduction of child chest radiation dose in digital radiography
GUO Hui,WU Hai-zhong,ZHANG Peng,ZENG Qun-li,ZHOU Xiao-shan,HE Xiao-ye,LI Bai-yan. Exploration of the feasibility on reduction of child chest radiation dose in digital radiography[J]. Journal of Xinjiang Medical University, 2011, 34(11): 1254-1257
Authors:GUO Hui  WU Hai-zhong  ZHANG Peng  ZENG Qun-li  ZHOU Xiao-shan  HE Xiao-ye  LI Bai-yan
Affiliation:GUO Hui,WU Hai-zhong,ZHANG Peng,ZENG Qun-li,ZHOU Xiao-shan,HE Xiao-ye,LI Bai-yan(Department of Radiology,First A f filiated Hospital,Xinjiang Medical University,Urumqi 830011,China)
Abstract:
Objective To explore the feasibility on reduction of radiation dose for child chest in digital radiography.Methods Children aged 0~14 were divided into five age groups,0~1,1~3,3~7,7~11,11~14 years old.900 cases,each age 180 cases,which were randomly divided into A,B,C groups.A group for Kodak DR 3000 fixed mAs group,B group for Kodak DR 3000 auto-mAs group,C group for Philips DR auto-mAs group.The mean mAs(radiation dose) and image quality among three groups were compared.Results(1) By compared mean mAs amo...
Keywords:radiation dose  image quality  children chest  digital radiography  