Self-refocused adiabatic pulse for spin echo imaging at 7 T
Authors:Balchandani Priti  Khalighi Mohammad Mehdi  Glover Gary  Pauly John  Spielman Daniel
Affiliation:Department of Radiology, Stanford University, Stanford, CA, USA. pritib@stanford.edu
Abstract:Spin echo pulse sequences are used to produce clinically important T(2) contrast. However, conventional 180° radiofrequency pulses required to generate a spin echo are highly susceptible to the B(1) inhomogeneity at high magnetic fields such as 7 Tesla (7 T), resulting in varying signal and contrast over the region of interest. Adiabatic 180° pulses may be used to replace conventional 180° pulses in spin echo sequences to provide greater immunity to the inhomogeneous B(1) field at 7 T. However, because the spectral profile of an adiabatic 180° pulse has nonlinear phase, pairs of these pulses are needed for proper refocusing, resulting in increased radiofrequency power deposition and long minimum echo times. We used the adiabatic Shinnar Le-Roux method to generate a matched-phase adiabatic 90°-180° pulse pair to obviate the need for a second adiabatic 180° pulse for phase refocusing. The pulse pair was then reformulated into a single self-refocused pulse to minimize the echo time, and phantom and in vivo experiments were performed to validate pulse performance. The self-refocused adiabatic pulse produced transmit profiles that were substantially more uniform than those achieved using a conventional spin echo sequence.
Keywords:adiabatic  Shinnar Le‐Roux  radiofrequency pulse  self‐refocused  matched‐phase  pulse‐pair  spin echo  B1 insensitive