Evaluation of a Noise Reduction Procedure for Chest Radiography
Authors:Ryohei Fukui  Rie Ishii  Kazuhiko Kodani  Yoshiko Kanasaki  Hisashi Suyama  Masanari Watanabe  Masaki Nakamoto  Yasushi Fukuoka
Affiliation:*Division of Clinical Radiology, Tottori University Hospital, Yonago 683-8504, Japan;†Division of Radiology, Matsue City Hospital, Matsue 690-8509, Japan;‡Division of Internal Medicine, Tottori Central Prefectural Hospital, Tottori 680-0901, Japan;§Division of Medical Oncology and Molecular Respirology, Tottori University Hospital, Yonago 683-8504, Japan
Abstract:

Background

The aim of this study was to evaluate the usefulness of noise reduction procedure (NRP),a function in the new image processing for chest radiography.

Methods

A CXDI-50G Portable Digital Radiography System (Canon) was used for X-ray detection. Image noise wasanalyzed with a noise power spectrum (NPS) and a burger phantom was used for evaluation of densityresolution. The usefulness of NRP was evaluated by chest phantom images and clinical chest radiography.We employed the Bureau of Radiological Health Method for scoring chest images while carrying out ourobservations.

Results

NPS through the use of NRP was improved compared with conventional image processing (CIP). The resultsin image quality showed high-density resolution through the use of NRP, so that chest radiographyexamination can be performed with a low dose of radiation. Scores were significantly higher than forCIP.

Conclusion

In this study, use of NRP led to a high evaluation in these so we are able to confirm the usefulness ofNRP for clinical chest radiography.
Keywords:chest radiography   image processing   noise power spectrum   noise reduction procedure   observation